Enhanced or Residual Oxygen MAP Gas Tester – 9000

For applications where a gas purge or air replacement process has been completed, a Bridge head space gas tester will confirm the residual oxygen level in the final package. Coupled to a gas sample line, a sample needle is used to penetrate the package exterior. While extracting a sample of the modified atmosphere, the analyzer will display an oxygen reading. After the test, the analyzer will freeze the display with the final reading. Subsequent tests are performed in a similar manner.

The MAP gas analyzer will test oxygen levels from above 0% to below 100%. If gas levels include CO2 levels above 20%, a special red band oxygen sensor can be used. The special sensor is more resistant to higher levels of CO2 than the standard oxygen sensor used in this oxygen analyzer. This tester can be configured to your specific application of various MAP packaging gas mixtures.

Applications include fresh cut produce, cosmetics, trail mix, snack packs and coffee packaging.

Enhanced or Residual Oxygen MAP Gas Analyzer

Down to 0.001% for Residual O2  |  Up to 100% for Enhanced Oxygen

Model 9000 MAP Head Space Gas Analyzer can assist in achieving and maintaining maximum quality control and extending shelf life of your food products.

  • Measures Residual Oxygen for Oxygen-Free MAP, or Oxygen fill for Enhanced Oxygen MAP.
  • Sealed for Harsh Environments – Industrial NEMA-4 Design.
  • Cordless Operation . Single or Double Shift Internal Battery.
    • Easy to use – Single button – Headspace O2 in 15 Seconds.
    • Customer Changeable Self-Calibrating E-C O2 sensor.
    • Small (6″x3″x3″), lightweight (2.5 Lbs), and rugged.

Model 900041 Gas Analyzer Specifications:

Gas Measured: Oxygen (O2)
Measurement Methods: Oxygen-Specific Electro Chemical Sensor – Auto-Calibrating.
Measuring Range: O2:    0.000-100.0%
Resolution: 0.001% Below 10.00%, 0.01% above 10.00%
Accuracy 4% relative to gas reading, or 0.010% (100ppm) absolute
Response time: 8 seconds to 95% Final Reading,
15 Seconds to 99% Final Reading.
Sampling Rate 4 ml/sec
Sample Time Factory default 15 Seconds. Adjustable in 1 second increments.
Gas-borne Contaminate Protection Built-In External and internal water blocking and particle filters protect analyzer from test gas-borne contamination.
Warm-up: Full accuracy in 30 seconds.
Operating Temperature: 0 to +40 degrees C.
Power:
Line Power
Internal Battery
10VAC / 220VAC 50/60 Hz Universal Wall Supply
8 hour NiMH battery pack. (18 hour pack optional).
O2 Sensor Life: 12-18 months
O2 Sensor Replacement: Customer Accessible – Simple 5 minute replacement with self-calibration – no need for factory return/service.
Weight:
Unit Alone
In Soft Case
2.5 Lbs.
3.5 Lbs.
Size:
Unit Alone
In Soft Case
6 ¼ L x 3W x 3 ½H
8 ½ L x 5W x 5H
PC Interface: Fully Sealed Serial Port, 9600 baud.

900041 MAP Gas Analzyer Brochure