CO/CO2/O2 – Model 900131 TriGas Analyzer

Headspace Case Ready Meat MAP Headspace TriGas Analyzer

900131

 

  • CO, CO2 and O2 Measurement Specifically for Case Ready Meat.
    • High Resolution 0.001% O2 & CO for Low-Ox.
    • 100.0% O2 and CO2 Range for High-Ox.
    • MAP Headspace Analyzer With Lo-Ox and Hi-Ox Capability.
  • Sealed for Line Environments — Industrial NEMA-4 Design.
  • Cordless Operation . 6 Hours (900131) or 16 hours (900132).
    • Fast and Easy to use . Test measurement in 15 Seconds.
    • Stable, Trouble-free Infrared CO2.
    • Unique Lo-Ox and Hi-Ox E-C O2 sensors for unmatched accuracy.
    • Compact (6″x3″x3″), lightweight, and rugged.

Model 900131/900132 Meat MAP Headspace Gas Analyzer Specifications:

Gases Measured: Carbon Monoxide (CO), Carbon Dioxide (CO2), and Oxygen (O2)
Measurement Methods: CO and CO2: NDIR (Non-Dispersive Infra-Red.) O2: Electro Chemical Sensor
Measuring Ranges: CO: 0.000% – 1.000%
CO2: 0.00% – 100.0%
O2: 0.00% – 100.0%
Resolution: CO: 0.001%, CO2: 0.01%,
O2: 0.01% above 2.00%, 0.001% Below 2.000%
Accuracy 4% relative to gas reading
Repeatability 1% relative to gas reading
Response time: 99% Final Reading in 15 Second.
Sampling Rate 4 ml/sec
Sample Time Factory default 15 Seconds. Adjustable in 1 second increments.
External Gas Contaminate Protection Pencil-point cross-drilled sample needle and customer serviceable particle/liquid blocking filter.
Internal Gas Contaminate Protection Internal particle/water blocking filter protects pneumatics and optical components from gas-borne contamination.
Warm-up: Full accuracy in 3 minutes.
Operating Temperature: 0 to +40 degrees C.
Power: Line Power
Internal Battery
110VAC / 220VAC Regulated Wall Supply
NiMH battery pack, 6-8 hours per charge . Model 900131.
(16 hour battery pack option available – Model 900132)
Weight:
Unit Alone
In Soft Case
2.5 Lbs.
3.5 Lbs.
Size:
Unit Alone
In Soft Case
6 ¼ L x 3W x 3 ½H
8 ½ L x 5W x 5H
PC Interface: Fully Sealed Serial Port, 9600 baud.

900131 Brochure